GasAlertMicroClip
Operator’s Manual
8
Pushbuttons
Pushbutton Description
C
To activate the detector, press C.
To deactivate the detector, press and hold C until the OFF countdown is complete and the LCD deactivates.
To view the TWA, STEL, and MAX (maximum) readings, press C twice. To clear the TWA, STEL, and MAX
readings, press C when the LCD displays RESET.
To initiate calibration, deactivate the detector. Press and hold C while the detector performs the OFF
countdown. Continue holding C while the LCD briefly deactivates and then begins the CAL countdown.
Release C when the CAL countdown is complete.
To activate the backlight while in normal operation, press C.
To acknowledge latched alarms, press C.
To acknowledge a low alarm and disable the audible alarm, press C (if the Low Alarm Acknowledge option
is enabled).