GasAlertMicroClip XT
Sensor Configuration
3. Click the Save to Device button located at the bottom of
the window.
4. The detector LCD automatically updates. The gas type
and sensor readings no longer display on the LCD for the
applicable sensor.
5. Enable the sensor as soon as possible. If the sensor is
damaged, replace it immediately. Refer to Replacing a
Sensor or Sensor Filter.
Calibration Gas Concentration
a Caution
The gas concentration value entered in Fleet Manager II must
match the gas concentration value on the gas cylinder.
1. Refer to the following list of recommended gas mixtures:
CO: 100 ppm balance N
2
•H
2
S: 25 ppm balance N
2
LEL: 50% LEL or 2.5 to 25% by vol. methane balance air
•O
2
: 18%
2. Select/enter the gas concentration value in the Cal Gas
field of the applicable sensor.
Calibration Interval
a Caution
BW recommends that the sensors be calibrated once every
180 days (6 months).
Define how often a sensor should be calibrated in the Cal Interval field.
A different calibration interval can be defined for each sensor.
1. Enter the value (0-365 days) for each sensor.
2. Enter 0 to disable the calibration interval option. Entering
zero automatically deactivates the Force Calibration
When Overdue user option.
The detector is shipped with the factory default set to 180 days.