GasAlertMicroClip
User Manual
Table 12. Troubleshooting Tips (cont.)
Problem Possible Cause Solution
Detector does not display normal
ambient gas reading after activation
self-test.
Sensor not stabilized
Detector requires calibration
Target gas is present
Used sensor: wait 60 seconds
New sensor: wait 5 minutes
Calibrate detector (see
Calibration)
Detector is operating properly. Use
caution in suspect areas
Detector does not respond to
pushbutton.
Battery is depleted
Detector is performing operations that
do not require user input
Charge battery (see
Charging the
Detector)
Pushbutton operation restored
automatically when the operation
ends
Detector does not accurately measure
gas.
Detector requires calibration
Detector is colder/hotter than
ambient gas
Sensor filter is blocked
Calibrate sensor (see
Calibration)
Allow the detector to acquire ambient
temperature before use
Clean the sensor filter (see
Replacing a Sensor or Sensor Filter)
Detector has been charging for over
3 hours and the charging indicator still
indicates that the battery has not
finished charging.
Battery is trickle charging Battery is full and ready for operation
40