GasAlertMicroClip
Troubleshooting
Table 12. Troubleshooting Tips (cont.)
Problem Possible Cause Solution
Detector does not enter alarm.
Alarm setpoint(s) are set incorrectly
Alarm setpoint(s) set to zero
Detector is in calibration mode
Reset alarm setpoints (see
Resetting
Gas Alarm Setpoints)
Reset alarm setpoints (see
Resetting
Gas Alarm Setpoints)
Complete the calibration procedure
Detector intermittently enters alarm
without apparent reason.
Ambient gas levels are near alarm
setpoint or the sensor is exposed to a
puff of the target gas
Alarms set incorrectly
Detector requires calibration
Missing or faulty sensor
Detector is operating normally. Use
caution in suspect areas. Check
maximum gas exposure reading.
Reset alarm setpoints (see
Resetting
Gas Alarm Setpoints)
Calibrate sensors (see
Calibration)
Replace sensor (see
Replacing a
Sensor or Sensor Filter)
Detector automatically deactivates.
Automatic shutdown feature activated
due to weak battery
Charge detector (see
Charging the
Detector)
Battery indicator does not display when
charging.
Detector may be depleted below
normal levels
Charge detector for approximately
8 hours. Detector LEDs may light
during first 5 hours. This is normal.
If the battery indicator does not light
after charging for 8 hours, contact
BW Technologies by Honeywell.
41